A AegiFlow
HIGHCVSS 8.8

CVE-2019-25719

CVE-2019-25719 updated by NVD

Modified
2026-07-24
Sources
nvd

Summary

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

References

Includes data from the National Vulnerability Database (NIST). NVD data is in the public domain; this page is not endorsed by NIST.

CVE® is a registered trademark of The MITRE Corporation. CVE content reproduced under the CVE Terms of Use; copyright designation © MITRE.